8-Ch Pulse Oximeter & CCD Spectrometer
Embedded C / ADS1299 Biopotential System & SolidWorks/STM32 Optical Baffle CAD
Designed and built an 8-channel ADS1299/Arduino biopotential pulse oximeter over SPI using Embedded C and Python, alongside a custom SolidWorks stray-light-baffled STM32 + Toshiba TCD1304 linear CCD spectrophotometer calibrated via 405/532/650 nm laser regression.
Optical Grating Angle
46.064° Incidence
Derived for 900 lines/mm grating to center m = 1 (532 nm) on CCD
Linear Detector Array
3,648 CCD Pixels
Toshiba TCD1304 (8 um pitch, 29.1 mm window) driven by STM32F401RE
Laser Calibration Fit
R² > 0.999 Linear
Three-point centroid regression across 405 nm, 532 nm, and 650 nm lasers
Biopotential Acquisition
8-Ch SPI ADS1299
Embedded C firmware & real-time Python physiological waveform UI
System Architecture & Quantitative Results
- Full-Stack 8-Channel Pulse Oximeter (Hardware & Software Lead): Interfaced an 8-channel
ADS1299biopotential ADC with an Arduino microcontroller over SPI using Embedded C firmware and a real-time Python physiological modelling and visualisation interface (building on a Year 1 I2C MaximMAX30101/MAX32664prototype). - SolidWorks Optical Enclosure (
Ahmed_Casing.SLDASM, 79% Report Mark): Modelled a light-tight 3D-printed housing integrating ThorLabs/Newport mounts, a 900 lines/mm transmission diffraction grating ( incidence angle), an asymmetrical sawtooth ceiling baffle, and a concave zero-order V-trap. - STM32 + TCD1304 Linear CCD Calibration: Calibrated the 3,648-pixel Toshiba
TCD1304detector in Python (scipy.stats.linregress) against , , and laser peaks to quantify transmission absorbance () and fluorescence.
SolidWorks 3D Optical CAD & Calibrated Spectral Results
[Fig. 1]Three-Laser Wavelength Calibration (405 / 532 / 650 nm) & Chemical Identification Proof
Left: Linear pixel-to-wavelength regression across violet (405 nm), green (532 nm), and red (650 nm) laser peaks. Right: Quantitative spectral matching of unknown chromophores and fluorophores against commercial reference baselines.


[Fig. 2]Normalised Transmission Absorbance & 90-Degree Fluorescence Spectra (n = 5)
Multi-replicate absorbance and orthogonal fluorescence emission curves with ±1 SD bands captured on the custom 3D-printed spectrophotometer.


[Fig. 3]SolidWorks Stray-Light Baffle Architecture: Sawtooth Ceiling & 45° Zero-Order V-Trap
Internal CAD cross-sections showing the asymmetrical sawtooth lid baffle teeth and the concave 45-degree specular V-trap engineered to extinguish zero-order (m = 0) and grazing reflections.


Want the Full Development Story?