Why Biopotentials Are So Hard to Clean Up: sEMG Conduction Velocity, EEG Alpha & Z-Plane ECG Filters

Breaking through integer-sample delay quantization to track 4.1 m/s muscle action potentials, measuring the 45 ms electromechanical delay between EMG and torque, taming EEG periodogram variance with Welch windows, and placing Z-plane poles and zeros by hand.

Ahmed Ahmed

Ahmed Ahmed

MEng Molecular Bioengineering, Imperial College London

MATLABDigital Signal ProcessingEMG / EEG / ECGWelch PSDZ-Transform

Why Biopotentials Are the Hardest Signals to Clean Up

When you first study digital signal processing on textbook synthetic sine waves, filtering feels almost trivial: take an FFT, pick a cutoff frequency, and multiply. The moment you connect actual electrodes to a human being, however, every assumption of stationary linear time-invariant (LTI) signal processing gets tested at once.

Across four biomedical signal processing modules at Imperial College London, I worked through the full hierarchy of clinical surface recordings in MATLAB: high-density surface electromyography (sEMG) on contracting skeletal muscle, occipital electroencephalography (EEG) during visual cortex Alpha blocking, and electrocardiography (ECG) corrupted by respiration baseline wander and 50 Hz50\text{ Hz} power-line interference. This article brings together the story of how I built each signal processing pipeline, why naive textbook estimators failed on the raw recordings, and how those lessons later fed directly into the DSP engine I wrote for MYOWAVE.

Biopotential Pulse Acquisition and Digital Storage Oscilloscope Bench Setup
Benchtop biopotential signal acquisition and digital oscilloscope verification: before any digital Z-plane filter or spectral estimator can extract a biomarker, the analog front-end and sampling rate must preserve phase and transient pulse morphology.

1. Tracking an Action Potential Moving at 4 Metres per Second (sEMG Delay & MFCV)

Our first problem looked deceptively simple: given two bipolar surface EMG channels x1[n]x_1[n] and x2[n]x_2[n] spaced a known distance Δd\Delta d apart along a muscle belly (moving away from the neuromuscular innervation zone toward the tendon), how fast is the depolarisation wave travelling along the muscle fibres?

Clinically, Muscle Fibre Conduction Velocity (MFCV=Δd/τ0\text{MFCV} = \Delta d / \tau_0) is a sensitive biomarker for peripheral neuropathy, muscular dystrophy, and localised muscle fatigue (where lactic acid accumulation slows sarcolemmal sodium-channel kinetics). If signal x2(t)x_2(t) is simply a delayed, noisy copy of x1(t)x_1(t) shifted by propagation time τ0\tau_0:

x2(t)=x1(t−τ0)+w(t)x_2(t) = x_1(t - \tau_0) + w(t)

then in discrete time we want to find the sample shift mm that minimises the Mean Squared Error (MSE\text{MSE}) between the two waveforms:

MSE(m)=1N∑n=0N−1(x2[n]−x1[n−m])2=P^x1+P^x2−2R^x1x2[m]\text{MSE}(m) = \frac{1}{N}\sum_{n=0}^{N-1}\big(x_2[n] - x_1[n - m]\big)^2 = \hat{P}_{x_1} + \hat{P}_{x_2} - 2\hat{R}_{x_1 x_2}[m]

Writing out that algebraic expansion in MATLAB was a neat "aha" moment: minimising the squared amplitude difference MSE(m)\text{MSE}(m) is mathematically identical to maximising the discrete cross-correlation R^x1x2[m]\hat{R}_{x_1 x_2}[m] when the channel powers are stationary.

Phase-Shifted Bipolar sEMG Waveforms, MSE Minimisation, and Delay-Aligned Overlay
Estimating muscle fibre propagation delay across a linear sEMG electrode array: raw phase-delayed motor-unit action potential trains (top), parabolic MSE minimisation (middle), and phase-aligned overlay after delay compensation (bottom).

The Sub-Sample Quantisation Trap

Here is the catch we ran into when we actually plugged in physical distances: if two electrodes sit Δd=10 mm\Delta d = 10\text{ mm} apart and action potentials travel along skeletal muscle at roughly 4.1 m/s4.1\text{ m/s}, the true propagation delay is only τ0≈2.44 ms\tau_0 \approx 2.44\text{ ms}. At a 1,000 Hz1,000\text{ Hz} sampling rate (Δt=1.0 ms\Delta t = 1.0\text{ ms}), an integer lag search can only return m=2 samplesm = 2\text{ samples} (5.0 m/s5.0\text{ m/s}) or m=3 samplesm = 3\text{ samples} (3.33 m/s3.33\text{ m/s}), which is a huge ±20%\pm 20\% velocity quantization error!

Mean Squared Error Objective Landscape Across Candidate Sample Lags
Zooming into the MSE bowl around the optimal sample delay: fitting a continuous local parabola across the three minimum discrete lag samples recovers sub-sample delay accuracy and a physiological conduction velocity of 4.12 m/s.

To break through the integer-sample ceiling without upsampling the entire recording by 20×20\times, we can either fit a three-point vertex parabola across the discrete MSE\text{MSE} minimum (m∗−1,m∗,m∗+1m^*-1, m^*, m^*+1) or move into the Fourier domain using the phase-shift property X2(ω)=X1(ω)e−jωτ0X_2(\omega) = X_1(\omega)e^{-j\omega\tau_0}, where τ0\tau_0 is the continuous linear slope of the cross-spectral phase angle ∠Sx1x2(ω)=−ωτ0\angle S_{x_1 x_2}(\omega) = -\omega\tau_0.


2. Why Your Muscle Lags Behind Your Nervous System (EMG Envelopes vs Isometric Torque)

In prosthetic control and biomechanics, you want to predict how much mechanical force a muscle is generating simply by reading its surface EMG voltage. However, a raw sEMG trace is a zero-mean bipolar interference pattern, if you average the raw voltage over a window, positive and negative phases cancel out to zero regardless of how hard the subject is contracting!

To extract an amplitude envelope proportional to motor-unit recruitment and rate coding, we full-wave rectified the signal (∣x[n]∣|x[n]|) and passed it through zero-phase low-pass Butterworth filters across progressively tighter cutoff frequencies (fc∈{1.5,3.0,6.0} Hzf_c \in \{1.5, 3.0, 6.0\}\text{ Hz}):

Rectified Low-Pass sEMG Envelopes Compared Against Load-Cell Isometric Joint Torque
Comparing full-wave rectified low-pass Butterworth sEMG envelopes against actual load-cell isometric joint torque: notice the bandwidth trade-off between ripple suppression and responsiveness, as well as the physiological electromechanical delay (EMD).

Overlaying our rectified sEMG envelope against the physical load-cell torque curve revealed two fascinating physiological facts:

  1. The Bandwidth-Latency Trade-off in Myoelectric Prosthetics: A 6.0 Hz6.0\text{ Hz} envelope reacts rapidly when the subject starts contracting, but leaves spiky motor-unit firing ripples that would make a prosthetic hand jitter. Dropping the cutoff to 1.5 Hz1.5\text{ Hz} produces a silky-smooth force estimate, but if you use a causal single-pass filter in real time, the filter group delay makes the prosthetic feel sluggish.
  2. Electromechanical Delay (EMD≈40 to 55 ms\text{EMD} \approx 40\text{ to }55\text{ ms}): Even when using a zero-phase (filtfilt) envelope, the electrical sEMG spike consistently rises roughly 45 ms45\text{ ms} before the mechanical torque transducer registers tension! That gap is the real physical time required for action potentials to travel down T-tubules, trigger dihydropyridine/ryanodine Ca2+\text{Ca}^{2+} release from the sarcoplasmic reticulum, cycle acto-myosin cross-bridges, and pull the slack out of the series elastic tendon.

3. Taming Spectral Variance in Cortical EEG Alpha Rhythms (Welch PSD)

In our third study, we examined occipital EEG recordings captured while a subject sat quietly with their eyes closed versus eyes open. When visual input is removed, thalamocortical feedback loops synchronise millions of pyramidal neurons in the primary visual cortex into a coherent 8 to 13 Hz8\text{ to }13\text{ Hz} Alpha oscillation (the Berger effect).

When I first computed a standard raw FFT periodogram S^PER(f)=1Nfs∣∑n=0N−1x[n]e−j2πfn/fs∣2\hat{S}_{\text{PER}}(f) = \frac{1}{N f_s}\left|\sum_{n=0}^{N-1} x[n]e^{-j2\pi fn/f_s}\right|^2 across the recording, the spectrum looked like a forest of jagged noise spikes. The reason is a classic statistical pitfall: the raw periodogram is an inconsistent estimator. Because each frequency bin is the sum of squared Gaussian random variables (a χ22\chi_2^2 distribution), the standard deviation of the periodogram at every frequency is equal to 100%100\% of the mean power itself, and taking a longer recording only packs more noisy frequency bins closer together without reducing the vertical variance!

Welch Averaged Modified Periodogram of Occipital EEG Alpha Band Synchronisation
Welch Power Spectral Density (PSD) of occipital EEG using 50%-overlapping Hann-windowed segments: variance reduction cleanly isolates the ~10.2 Hz cortical Alpha peak from the 1/f background neural spectrum.

Switching to Welch's averaged modified periodogram, slicing the EEG trace into KK overlapping segments of length LL, tapering each segment with a smooth Hann window w[n]w[n] to suppress sidelobe spectral leakage, and averaging their periodograms, cuts the estimator variance by roughly 911K\frac{9}{11 K}:

S^Welch(f)=1KLUfs∑k=0K−1∣∑n=0L−1w[n] xk[n] e−j2πfn/fs∣2,U=1L∑n=0L−1w2[n]\hat{S}_{\text{Welch}}(f) = \frac{1}{K L U f_s}\sum_{k=0}^{K-1}\left|\sum_{n=0}^{L-1} w[n]\,x_k[n]\,e^{-j2\pi fn/f_s}\right|^2, \quad U = \frac{1}{L}\sum_{n=0}^{L-1} w^2[n]

With the variance suppressed, the ∼10.2 Hz\sim 10.2\text{ Hz} occipital Alpha peak stood out unmistakably above the 1/f1/f cortical background noise.


4. Placing Poles and Zeros by Hand on the Complex Z-Plane (ECG Filtering)

In our final study, we were handed a clinical ECG trace that was almost unreadable: heavy 50 Hz50\text{ Hz} mains hum blurred the baseline into a thick ribbon, while slow 0.2 Hz0.2\text{ Hz} respiratory motion swung the trace up and down across several millivolts.

Rather than calling a black-box filter wizard, we designed a custom Infinite Impulse Response (IIR) pole-zero filter directly on the complex zz-plane. The intuition behind pole-zero placement is wonderfully geometric when you view the frequency response H(ejω)H(e^{j\omega}) as the ratio of distances from a point ejωe^{j\omega} moving around the unit circle to the zeros zkz_k and poles pkp_k:

  • To completely extinguish 50 Hz50\text{ Hz} interference (and DC drift at ω=0\omega = 0), you drop conjugate transmission zeros directly onto the unit circle at angle ω0=±2π(50 Hz)/fs\omega_0 = \pm 2\pi(50\text{ Hz})/f_s (and z=1z = 1), forcing the numerator of H(ejω0)H(e^{j\omega_0}) to vanish to zero.
  • If you only place zeros, however, the trench around ω0\omega_0 is wide and dull, chewing away legitimate 25 to 45 Hz25\text{ to }45\text{ Hz} QRS complex energy. To make the notch razor-sharp while keeping the system causally BIBO-stable, you place conjugate poles just inside the unit circle along the same radial angle at radius r=0.96r = 0.96 (p1,2=re±jω0p_{1,2} = r e^{\pm j\omega_0}, with ∣p∣<1|p| < 1):

Hnotch(z)=b0(1−ejω0z−1)(1−e−jω0z−1)(1−rejω0z−1)(1−re−jω0z−1)=b01−2cos⁡(ω0)z−1+z−21−2rcos⁡(ω0)z−1+r2z−2H_{\text{notch}}(z) = b_0 \frac{\left(1 - e^{j\omega_0}z^{-1}\right)\left(1 - e^{-j\omega_0}z^{-1}\right)}{\left(1 - r e^{j\omega_0}z^{-1}\right)\left(1 - r e^{-j\omega_0}z^{-1}\right)} = b_0 \frac{1 - 2\cos(\omega_0)z^{-1} + z^{-2}}{1 - 2r\cos(\omega_0)z^{-1} + r^2 z^{-2}}

Complex Z-Plane Pole-Zero Diagram of Our IIR ECG Filter
Complex z-plane pole-zero plot verifying unit-circle transmission zeros and interior stabilizing poles (|p| < 1).
Restored P-QRS-T ECG Morphology and Filter Frequency Response
Time- and frequency-domain validation showing clean recovery of the P wave, QRS complex, and ST-T segment.

Limitations and Real-World Clinical Considerations

Working across all four biosignal modalities highlighted why naive digital filters can actually cause misdiagnoses if applied carelessly:

  1. Causal IIR Phase Distortion vs ST-Segment Elevation: Near a high-Q pole-zero notch or a steep high-pass cutoff (0.5 Hz0.5\text{ Hz}), the phase response ∠H(ejω)\angle H(e^{j\omega}) of a causal single-pass IIR filter becomes strongly non-linear. On an ECG, non-linear phase delay shifts the low-frequency T wave relative to the high-frequency QRS complex, creating artificial ST-segment depression or elevation that mimics myocardial ischaemia! In offline analysis (and buffered sliding windows), running zero-phase forward-backward filtering (filtfilt) cancels the phase angle identically (∠∣H(ejω)∣2=0\angle |H(e^{j\omega})|^2 = 0), preserving clinical ST morphology.
  2. Non-Stationarity During Muscle Fatigue and Seizures: Welch's PSD assumes the signal is wide-sense stationary over the averaging window. When a muscle fatigues during a sustained contraction (shifting median sEMG power toward lower frequencies) or when a cortical EEG transitions suddenly into a transient spike-and-wave burst, time-averaging smears the transition. Extending these pipelines with Continuous Wavelet Transforms (CWT, Morlet wavelets) or Short-Time Fourier Transform (STFT) spectrograms captures joint time-frequency evolution without sacrificing transient localisation.